• Only connect to a grounded socket. If the power supply outlet is not grounded, it will be necessary to

have a ground installed by qualified personnel.

• It is of utmost importance to set up a password during the first start up procedure. Make sure that the

password is available to the person responsible.

2.1.2  Cautions

• Ensure that there is sufficient room around the unit for air circulation. A good air circulation especially

around the compressor is essential for trouble-free operation. Keep the ventilation openings of the com-
pressor cover always free from any obstruction. Make sure that a minimum distance is provided between
the unit and any wall or other device located next to it.

• Do not locate the unit below a ceiling fan or right next to air-conditioning equipment.
• The sound intensity level produced by the appliance is lower than 70 dBA  (measured at a distance of

1 meter).

• Check whether the appliance has been delivered undamaged. If you find that damage has occurred in

transit, immediately contact the delivery service or relevant sales outlet, submitting the delivery note or
proof of purchase. 

• Do not operate a unit that has been damaged in transit! If you are unsure, contact your sales outlet and

ask them.

• The device must be transported in an upright position only (maximum inclination 45°).
• All connected devices/signalling equipment must provide reinforced or double insulation for protection

against electric shock.

• Upon the occurrence of an alarm situation, attempts must be made to discover the reason for the alarm

and resolve it as quickly as possible. If that is not successful the necessary measures must immediate-
ly be taken so that the stored goods are not damaged.

• Optimal function of the appliance is given at an ambient temperature indicated in the technical data (see

chapter 11) with a maximum relative humidity of 95%.

• The device must be set up in a dry well-ventilated place. Avoid direct sunlight or locating it close to a

heat source. The location must be protected against rain and dust.

• This device cannot be used/is not for use in facilities at altitudes of 2000m or higher.
• The use of hydrogen peroxide H2O2 (VHP) for the disinfection (sterilization) of the appliances is not suit-

able. This procedure may be applied exclusively with dedicated appliances.

• Other than the cleaing procedure described in this manual, follow the internal policy and procedures.
• Before using the refrigerator it should be cleaned inside and outside.
• For cleaning, use only gentle cleaning agents. Never use aggressive or caustic cleaning agents, scour-

ing powder, steel wool, abrasive sponges or chemical solvents. When cleaning, make sure that no flu-
ids of any kind run into the ventilation housing.

• For disinfecting, we recommend all the surface disinfecting agents commonly used by the customer, pro-

vided they are recommended by the national organizations. For disinfecting small areas, we recommend
using a concentrated alcoholic agent.

• Before being loaded, the unit must have reached the set point temperature.
• The load should only be stored in the provided baskets and should not exceed the storage areas. The

total amount of stored blood bags should be evenly distributed on the 5 baskets (max. 18 blood bags
per basket) (image 6). Do not overload the device.

• Different blood or plasma types should be sorted and stored in a manner for the User to identify these

with easy. 

• The loading must never obstruct the air-inlet-openings.
• Do not store warm items in the devices. They are not designed for fast cooling of warm items. 
• Avoid long and / or frequent lid openings to prevent that the inner temperature rises to high.
• Make sure that the lid is correctly closed by using the supplied latches and that the lid gasket seals cor-

rectly against the cabinet. 

Summary of Contents for MB 3000 G

Page 1: ...820 9505 61 ed0821 Instructions for Use MB 3000 G ICE LINED BLOOD STORAGE REFRIGERATOR...

Page 2: ...2...

Page 3: ...3...

Page 4: ...rtified according to the safety standard IEC 60335 1 This equipment is designed to be safe at least under the following condi tions based on the IEC 60335 1 Indoor use Altitude up to 2000 m Temperatur...

Page 5: ...Storage_______________________________________18 4 4 Control Panel ____________________________________________________19 4 4 1 Overview ________________________________________________19 4 4 2 Displa...

Page 6: ...6 1 Model Overview Group Model name Ice lined Blood storage Refrigerators MB3000G...

Page 7: ...ce lining consists of icepacks with water freezing during its operation During periods of power failure the ice bank acts as cold storage to protect the goods stored in the MB3000G The unit keeps the...

Page 8: ...requency other than that on the nameplate may cause fire or electric shock Do not store flammable or explosive substances inside these appliances e g aerosols with flammable propellant These may cause...

Page 9: ...he technical data see chapter 11 with a maximum relative humidity of 95 The device must be set up in a dry well ventilated place Avoid direct sunlight or locating it close to a heat source The locatio...

Page 10: ...ubmitting the delivery note or proof of purchase Do not operate a unit that has been damaged in transit If you are unsure contact your sales outlet and ask them The unit must be transported in an upri...

Page 11: ...e indoors in hospitals blood donation services blood transfusion centers and clinical laboratories The devices are designed for an ambient temperature range from 10 C to 43 C under operation and from...

Page 12: ...od or blood components The devices are not intended to be used for the cooling of foodstuffs or any other products not covered in the intended purpose The devices are not intended to be operated outsi...

Page 13: ...ectric shock may result due to corrosion The wall socket must be easily accessible When the unit is in its final position it is important to immobilise the unit using the feet CAUTION Optimal function...

Page 14: ...refrigerator it should be cleaned inside and outside For cleaning use only gentle cleaning agents Never use aggressive or caustic cleaning agents scour ing powder steel wool abrasive sponges or chemi...

Page 15: ...s may cause electric shock When disconnecting always pull the plug and never the cable Pulling the cord may result in electric shock or fire by short circuit CAUTION Before connecting and switching th...

Page 16: ...of the cabinet into the lining compartment between the inner and outer cabinet Watch out to introduce the holders in the correct way and not to damage the evapo rator tubes To ensure the best transpo...

Page 17: ...that the unit may be locked using a key Always ensure that the key is available to the user It is of utmost importance to set up a password during the first start up procedure Make sure that the passw...

Page 18: ...s Do not store warm items in the devices They are not designed for fast cooling of warm items Avoid long and or frequent lid openings to prevent that the inner temperature rises to high Make sure that...

Page 19: ...e disabled through the settings The temperature values are displayed with a resolution of 0 1 Refer to chapter 6 4 for more informa tion 4 5 Temperature Display During normal operation the display sho...

Page 20: ...MENU key q select EXTRA 8 q LANGUAGE 8 q select your language 8 5 3 The SETTINGS Menu The parameters contained in the SETTINGS menu partially intervene in the function of the appliance Access to the...

Page 21: ...nction q HIDE SERVICE WARNING 8 This function is for cancelling the service message q USER PASSWORD 8 Changing the password to access the SETTINGS MENU w CHANGE USER PASSWORD 0XXX XXXX 8 Enter the new...

Page 22: ...pressing the button during a power failure 100 approx 10 Sec 50 approx 5 Sec see chapter Alarm situations q VOLTAGE IND 8 w LINE VOLTAGE pq NONE 8 Shows the input voltage of the appliance on the dis...

Page 23: ...is connected to the appliance and the display in the EXTRA menu activated 3 EXTRA AMBIENT PROBE this message appears followed by the currently measured ambient temperature in the display w REG C If th...

Page 24: ...interrupted with the appliance switched on A power failure alarm is stored in the alarm history The alarm mes sage disappears as soon as the power supply is restored If one of the following error mess...

Page 25: ...the corresponding alarm message appears on the display 6 3 Alarm History The alarm history list contains all relevant data on temperature and powerfail events The list contains up to 20 alarm situatio...

Page 26: ...disposed of separately in the case of a fault 7 2 Remote contact There are two terminals with three floating contacts on the back of the unit that can be used to trigger an additional external alarm...

Page 27: ...de and outside For cleaning use only gentle cleaning agents Never use aggressive or caustic cleaning agents scour ing powder steel wool abrasive sponges or chemical solvents When cleaning make sure th...

Page 28: ...cabinet frame Close the lid Connect the power supply plug to the wall socket and switch on the MB3000G Allow it to run and observe the cabinet temperature Put back the contents into the MB3000G from...

Page 29: ...lly read and follow the information and advice below Problem Interior temperature too hot Please check if large quantities of warm goods have been placed inside the unit causing a surge of the interio...

Page 30: ...cold alarm C Preset warm alarm C Ambient operating temperature range C Storage and transport temperature not operational C Max Humididty range Altitude Refrigerant Type Refrigerant quantity g Hold ove...

Page 31: ...31...

Page 32: ...B Medical Systemes S r l op der Hei 17 L 9809 Hosingen Luxembourg...

Reviews: