9
• Only connect to a grounded socket. If the power supply outlet is not grounded, it will be necessary to
have a ground installed by qualified personnel.
• It is of utmost importance to set up a password during the first start up procedure. Make sure that the
password is available to the person responsible.
2.1.2 Cautions
• Ensure that there is sufficient room around the unit for air circulation. A good air circulation especially
around the compressor is essential for trouble-free operation. Keep the ventilation openings of the com-
pressor cover always free from any obstruction. Make sure that a minimum distance is provided between
the unit and any wall or other device located next to it.
• Do not locate the unit below a ceiling fan or right next to air-conditioning equipment.
• The sound intensity level produced by the appliance is lower than 70 dBA (measured at a distance of
1 meter).
• Check whether the appliance has been delivered undamaged. If you find that damage has occurred in
transit, immediately contact the delivery service or relevant sales outlet, submitting the delivery note or
proof of purchase.
• Do not operate a unit that has been damaged in transit! If you are unsure, contact your sales outlet and
ask them.
• The device must be transported in an upright position only (maximum inclination 45°).
• All connected devices/signalling equipment must provide reinforced or double insulation for protection
against electric shock.
• Upon the occurrence of an alarm situation, attempts must be made to discover the reason for the alarm
and resolve it as quickly as possible. If that is not successful the necessary measures must immediate-
ly be taken so that the stored goods are not damaged.
• Optimal function of the appliance is given at an ambient temperature indicated in the technical data (see
chapter 11) with a maximum relative humidity of 95%.
• The device must be set up in a dry well-ventilated place. Avoid direct sunlight or locating it close to a
heat source. The location must be protected against rain and dust.
• This device cannot be used/is not for use in facilities at altitudes of 2000m or higher.
• The use of hydrogen peroxide H2O2 (VHP) for the disinfection (sterilization) of the appliances is not suit-
able. This procedure may be applied exclusively with dedicated appliances.
• Other than the cleaing procedure described in this manual, follow the internal policy and procedures.
• Before using the refrigerator it should be cleaned inside and outside.
• For cleaning, use only gentle cleaning agents. Never use aggressive or caustic cleaning agents, scour-
ing powder, steel wool, abrasive sponges or chemical solvents. When cleaning, make sure that no flu-
ids of any kind run into the ventilation housing.
• For disinfecting, we recommend all the surface disinfecting agents commonly used by the customer, pro-
vided they are recommended by the national organizations. For disinfecting small areas, we recommend
using a concentrated alcoholic agent.
• Before being loaded, the unit must have reached the set point temperature.
• The load should only be stored in the provided baskets and should not exceed the storage areas. The
total amount of stored blood bags should be evenly distributed on the 5 baskets (max. 18 blood bags
per basket) (image 6). Do not overload the device.
• Different blood or plasma types should be sorted and stored in a manner for the User to identify these
with easy.
• The loading must never obstruct the air-inlet-openings.
• Do not store warm items in the devices. They are not designed for fast cooling of warm items.
• Avoid long and / or frequent lid openings to prevent that the inner temperature rises to high.
• Make sure that the lid is correctly closed by using the supplied latches and that the lid gasket seals cor-
rectly against the cabinet.
Summary of Contents for MB 3000 G
Page 1: ...820 9505 61 ed0821 Instructions for Use MB 3000 G ICE LINED BLOOD STORAGE REFRIGERATOR...
Page 2: ...2...
Page 3: ...3...
Page 6: ...6 1 Model Overview Group Model name Ice lined Blood storage Refrigerators MB3000G...
Page 31: ...31...
Page 32: ...B Medical Systemes S r l op der Hei 17 L 9809 Hosingen Luxembourg...