Trigger
63
Slope Trigger
The slope trigger looks for a rising or falling transition from one level to another level in greater than or
less than a certain amount of time.
In the oscilloscope, positive slope time is defined as the time difference between the two crossing points
of trigger level line A and B with the positive edge as shown in the figure below.
Figure 5.11
Slope Trigger
Coupling and noise reject can be set in edge trigger, see the sections
, and
.
To set the slope of the trigger:
1. Press the
button to enter the trigger menu.
2. Press the
button to select
Slope
.
• A dropdown menu containing the available options will appear.
Figure 5.12
Slope
3. Use the
to navigate the available options.
• Repeately pressing
will toggle between
Rising
and
Falling
.