Meter
166
17.2 DCV/ACV
To enable DCV or ACV metering and set the range:
Step 1.
Press the
key.
Step 2.
Press the
softkey to view the available metering option.
• Use the
to select
"DCV
or
"ACV"
.
Step 3.
Press the
softkey to view the available mode options.
• Use the
to one of the available modes.
• If manual was selected press the
softkey to toggle between the available ranges.
When DCV or ACV is applied the following ranges are available:
DCV/ACV Metering
Function
Range
Resolution
Accuracy
DC Voltage
60.00 mV
10
μ
V
(
±
1%
±
15 digit)
600.00 mV
100
μ
V
(
±
1%
±
5 digit)
6.00 V
1 mV
60.00 V
10 mV
600.00 V
100 mV
1000.00 V
1 V
(
±
1.5%
±
15 digit)
AC Voltage
45 hz to 400Hz
60.00 mV
10
μ
V
(
±
1%
±
15 digit
600.00 mV
100
μ
V
(
±
1%
±
5 digit
6.00 V
1 mV
60.00 V
10 mV
600.00 V
100 mV
750.00 V
1 V
(
±
1.5%
±
15 digit)
Table 17.2
DCV & ACV Ranges