System Settings
148
13.2 Self Cal
The self-calibration program can quicckly make the oscilloscope reach the best working state to get the
most precise measurement values. You can perform the self-calibration at any time, especially when
the environment temperature change is up to or more than 5
∘
C. Ensure that the oscillosope has been
warmed up or operated for more than 30 minutes before performing a self-calibration.
To perform a self-calibration:
Step 1.
Disconnect all the input channels.
Step 2.
Press the
key to enter the
SETTINGS
menu.
Step 3.
Press the
softkey to navigate to page
2/3
of the
SETTINGS
menu.
Step 4.
Press the
softkey to enter the self-calibration notice.
Figure 13.2
Do Self Cal
Step 5.
Press the
softkey to begin the self-calibration.
• During calibration most of the keys are disabled.
• Once the self-calibration is complete, the message
"Press Run/Stop to exit"
will appear.
Step 6.
Press the
button to exit the calibration interface.