Perfusor® compact, 2.1 gb
B - 1
B
B -
Modification Instructions for Syringe
Holder
Description
The syringe holder must be modified as recognition of syringes
with a large outer diameter by the light barrier system was diffi-
cult. This modification is limited to installation of a washer.
Modification
1.
Pierce through the cap and remove.
2.
Fasten syringe holder with pin punch.
3.
Remove screw.
4.
Pull off holder.
5.
Push washer 0.8x4x3.2 on the shaft.
6.
Insert syringe holder.
7.
Fit new screw (not the old one) and safety lock with Loctite
274.
8.
Place on cap.
9.
Check syringe recognition for proper function
Fig.: B - 1
B - 2
Summary of Contents for Perfusor compact
Page 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10...
Page 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes...
Page 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes...
Page 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes...
Page 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes...
Page 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes...
Page 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes...
Page 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb...
Page 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes...
Page 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes...
Page 76: ...10 4 Perfusor compact 2 1 gb For your notes...
Page 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes...
Page 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes...