Unit Elements
4
Perfusor® compact, 2.1 gb
4 - 17
4.17 Housing Bottom Part, Complete
Designation
Ord. No.
Housing bottom part, complete . . . . . . . . . . . . . . . . . . 3450 6594
Exchange
1.
Open housing
2.
Shift type plate.
a)
Warm up type plate with a hair dryer until the adhesive
can be removed (not too hot as otherwise the housing is
damaged).
b)
Clean adhesive position on new housing and stick type
plate. New type plates can only be ordered as spare parts
if the old type plates are returned to B.Braun
ring of Spare Parts and Test Equipment“
3.
Remove drive from old housing
4.
Install drive in new housing
5.
Close housing.
Note
Do not squeeze the cable
6.
Calibrate in Service Program
Summary of Contents for Perfusor compact
Page 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10...
Page 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes...
Page 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes...
Page 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes...
Page 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes...
Page 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes...
Page 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes...
Page 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb...
Page 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes...
Page 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes...
Page 76: ...10 4 Perfusor compact 2 1 gb For your notes...
Page 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes...
Page 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes...