Unit Elements
4
4 - 6
Perfusor® compact, 2.1 gb
4.3
Syringe Holder
Designation
Ord. No.
Syringe holder, complete . . . . . . . . . . . . . . . . . . . . . . . 3450 6608
with screw and cap
Exchange
1.
Pierce through the cap and remove.
2.
Fasten syringe holder with pin punch.
3.
Remove screw.
4.
Pull off holder.
5.
Push spacer washer 0.8x4x3.2 on the shaft if the syringe is
not recognized.
6.
Insert new syringe holder.
7.
Fit new screw (not the old one) and safety lock with Loctite
274.
8.
Replace new cap.
4.4
Unit Feet
Designation
Ord. No.
Unit feet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 6640
Note
The feet can be turned and used once again. Pull feet out and turn
around or exchange.
Fig.: 4 - 5
Summary of Contents for Perfusor compact
Page 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10...
Page 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes...
Page 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes...
Page 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes...
Page 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes...
Page 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes...
Page 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes...
Page 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb...
Page 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes...
Page 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes...
Page 76: ...10 4 Perfusor compact 2 1 gb For your notes...
Page 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes...
Page 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes...