Unit Elements
4
Perfusor® compact, 2.1 gb
4 - 9
4.8
LS-Clip
Designation
Ord. No.
LS-clip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 7710
Exchange
1.
Open unit
2.
Press buzzer out of the holder.
3.
Pull LS-clip out of the guide and exchange.
4.
Assembly is done in reverse order.
Setting the Loudness of the Alarm Tone (from serial number
38100 on)
1.
Open battery compartment
2.
Remove batteries.
3.
Connect unit to mains and switch unit on.
4.
Disconnect from mains for a short moment when the switch-
on test is finished. Pull mains connector and plug in again to
trigger a device alarm (code 22, continuous tone).
5.
Put a small flat blade screw driver (carefully) through the
battery compartment opening and set the volume desired.
6.
Switch unit off via the keyboard.
7.
Insert batteries.
8.
Close battery compartment.
Fig.: 4 - 9
LS-clip
Fig.: 4 - 10
Summary of Contents for Perfusor compact
Page 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10...
Page 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes...
Page 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes...
Page 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes...
Page 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes...
Page 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes...
Page 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes...
Page 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb...
Page 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes...
Page 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes...
Page 76: ...10 4 Perfusor compact 2 1 gb For your notes...
Page 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes...
Page 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes...