Unit Elements
4
Perfusor® compact, 2.1 gb
4 - 13
8.
Pay attention to cable laying (please see
).
9.
Close unit
Note
Do not squeeze cable.
10. Calibrate in Service Program
(see „Calibration after Replace-
4.14 Axial Positioner
Designation
Ord. No.
Axial positioner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 6659
Exchange
1.
Open unit
2.
Move drive arm to middle position and lock.
3.
Loosen both screws on drive.
4.
Lift drive until the axial positioner is free.
5.
Remove axial positioner by forcing apart. Replace new axial
positioner and make sure that the scraper ring is correctly fit-
ted.
6.
Assembly is done in reverse order.
Note
Do not squeeze the cable
7.
Calibration is required in Service Program
, as the drive was dis-
mounted.
Fig.: 4 - 15
Summary of Contents for Perfusor compact
Page 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10...
Page 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes...
Page 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes...
Page 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes...
Page 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes...
Page 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes...
Page 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes...
Page 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb...
Page 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes...
Page 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes...
Page 76: ...10 4 Perfusor compact 2 1 gb For your notes...
Page 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes...
Page 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes...