Checks after Repair
Depending on the work carried out, perform the relevant check
blocks (1., 2. and / or 3.).
1. Visual Inspection
2. Electrical Safety
according to IEC / EN 60601-1
or VDE 0750 and VDE 0751
3. Functional Inspection
❒
Infusomat® fmS:
Cleanliness, completeness, damage and
faults affecting safety
Particularly:
❒
Damage to and readability of the
labels
❒
Completeness, damage and
readability of the additional labels
on an Infusomat® fmS set for the
Infusomat® Space Line
❒
Pump sealing diaphragm
❒
Membrane keyboard
❒
Rubber feet
❒
Operating unit
❒
Locking mechanism of the pump
cover
❒
Flow inhibitor
❒
Mains plug connector
❒
MFC connector incl. MFC cable
❒
Check voltage values
100/110/120 V = T 0.315 A
200/230/240 V = T 0.16 A
❒
Mains cable
❒
Drop sensor with cable and plug
connector
❒
Mains voltage
______ V AC
❒
Protective conductor
resistance incl. mains
cable < 0.2
_______
❒
Insulation resistance
>> 2 M
_________
❒
Earth leakage current at NC
30
A ________
A
Switch on unit
❒
Self-test
❒
Audible alarm
❒
Status displays
❒
LCD lighting
❒
Display (all symbols and pixels) in the
LCD display
Operation
❒
Opening and closing mechanism of the
door
❒
Functions of the buttons
❒
Enter delivery rate, volume and time.
The values displayed in the LCD display
correspond to the values entered.
❒
Staff call
Test incl. MFC staff call line, if used
❒
Pump cover alarm
❒
Alarm suppression
Pressure cut-off, electronic
(Reference values measured with
electronic pressure sensor)
Delivery rate: 50 ml/h
Volume: 250 ml
Test without drop sensor
Note
Depending on the setting, the measurement
should be performed only with one of the
two lines.
❒
With original Infusomat® line
❒
Pressure level, low
❒
Pressure level, middle
❒
Pressure level, high
❒
With Infusomat® Space line
❒
Pressure level, low
❒
Pressure level, middle
❒
Pressure level, high
(Part 1 of 2)
5 - 2
Summary of Contents for Infusomat fmS
Page 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10...
Page 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes...
Page 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes...
Page 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes...
Page 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes...
Page 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes...
Page 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes...
Page 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes...
Page 84: ...Index 11 11 2 Infusomat fmS 2 3 gb...
Page 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes...