Unit Elements
4
4 - 10
Infusomat® fmS 2.3 gb
Exchange Mains Module
1.
Loosen 3 screws and remove the power supply unit.
Note
The voltage selection for switchable power supplies is at the fuse
element.
Exchange Drop Sensor Socket
1.
Loosen nut and exchange drop sensor socket.
2.
Safety lock with Sicomet 50.
3.
Make ground connection.
Check
Safety check, functional check.
Fig.: 4 - 11
Power supply
3 X Torx M3x10
Summary of Contents for Infusomat fmS
Page 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10...
Page 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes...
Page 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes...
Page 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes...
Page 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes...
Page 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes...
Page 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes...
Page 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes...
Page 84: ...Index 11 11 2 Infusomat fmS 2 3 gb...
Page 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes...