Unit Elements
4
Infusomat® fmS 2.3 gb
4 - 15
c)
Push pressure sensor with pressure sensor board to the
rear until stopper.
d)
Move lower slide of the finger pump rearwards.
e)
Note down pressure sensor value.
f)
Insert 4 mm calibration gauge.
g)
Push pressure sensor with pressure sensor board slightly
forward.
h)
Tighten screw (2).
The measured value read on the pressure sensor must be
5 to 15 digits above the value noted.
Note
Pull nut slightly out of the pressure sensor housing.
i)
After calibration return to the initial function with END.
j)
Select SAVE NO. (Do not actuate the YES key.)
k)
Switch off unit.
7.
Assembly is done in reverse order.
8.
Check the user data and reenter, if necessary, as data was
saved in the EEPROM of the exchanged light barrier board.
9.
Calibrate pressure sensor:
Note
Calibration weights are required to calibrate the pressure sensor.
An alignment with infusion lines is not permitted.
a)
Clean the pump front side.
b)
Place the Infusomat® fmS without mains lead and drop
sensor and the service connector inserted horizontally
(front facing upwards) in the cellular packing of the cal-
ibration device.
c)
Switch on unit and call in the pressure sensor reading in
the Service Program
(see „Pressure Sensor Function
d)
Open pump cover.
e)
Fasten holders for the calibration weights under the
pump cover and let snap in at the locking bow.
f)
Position weight 1 (100 g) for 400 mbar carefully in the
holder on the pressure sensor. The weight must be placed
without any friction in the holder.
g)
Note down measured value.
h)
Repeat steps f) and g) with weight 2 (128 g) for 800 mbar
and weight 3 (166 g) for 1200 mbar.
Summary of Contents for Infusomat fmS
Page 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10...
Page 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes...
Page 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes...
Page 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes...
Page 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes...
Page 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes...
Page 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes...
Page 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes...
Page 84: ...Index 11 11 2 Infusomat fmS 2 3 gb...
Page 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes...