Software
2
Infusomat® fmS 2.3 gb
2 - 5
Control Microprocessor 68hc11
FFxx is displayed in the 7 segment display with flashing dots. FFxx
is the error code.
7 Segment Display Description
FF01
dummy for test
FF02
battery not present / missing battery current
FF03
defective RAM memory
FF04
defective program memory
FF05
defective program memory
FF06
calibration data error from EEPROM
FF07
pump head cycle not plausible
FF08
failure / inaccuracy of system clock
FF09
failure 100msec system clock
FF10
reset during active operation
FF12
no dynamic pressure sensor signal (EDB)
FF16
defective membrane keyboard (from IFMC02003 on)
Table 2 - 2