Service Program
3
3 - 8
Infusomat® fmS 2.3 gb
Menu
Function 450.0
The availability of menus on the user interface can be set.
-
Double rate entry
-
Online rate entry
-
Interval Bolus
1.
OK activates the function.
2.
Activate / deactivate the decimal function with NEXT.
3.
Acknowledge with YES.
4.
Return to main menu with END.
Delivery Rate
Function 460.0
The maximum and minimum delivery rates can be set. Range of
adjustment of the delivery rate: 0.1 to 999.9 ml/h
1.
OK activates the function.
2.
Select min./max. delivery rate with NEXT.
3.
Acknowledge with YES.
4.
Enter value with a numeric key.
5.
Acknowledge with OK.
6.
Return to initial function with END.
Air Alarm
Function 470.0
The air inline sensor sensitivity of the air rate alarm in ml/h (total
air alarm) and of the maximum air bubble in ml (single bubble)
can be adjusted.
Setting range air rate: 0.5 to 3.5 ml/h
Setting range air bubble: 0.01 to 0.3 ml/h
1.
OK activates the function.
2.
Select air rate (ml/h) or air bubble (ml) with NEXT.
3.
Acknowledge with YES.
4.
Enter value with a numeric key.
5.
Acknowledge with OK.
6.
Return to initial function with END.
Summary of Contents for Infusomat fmS
Page 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10...
Page 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes...
Page 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes...
Page 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes...
Page 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes...
Page 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes...
Page 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes...
Page 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes...
Page 84: ...Index 11 11 2 Infusomat fmS 2 3 gb...
Page 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes...