Service Program
3
Infusomat® fmS 2.3 gb
3 - 9
Dianet Mode Display
Function 480.0
When operated with DianetStar the respective DianetStar-mode
(CA, CC, CD) with address 01, e.g. mode CA and address 01 is dis-
played by: „###CA01###“.
The duration of the display after the last data transmission can be
set between 0 and 255 seconds.
Deleting Dose Data
Function 490.0
When this function is activated operation can be continued with
the dose data of the previous therapy.
Calibration
Note
All safety relevant parameters are set by the manufacturer. If
these parameters are changed, a new calibration must be per-
formed with calibrated test equipment.
Scale Factor
Function 510.0
The scale factor can be set in the limits 40 to 99 digits. Every digit
step is equivalent to a 0.5 % modification of the delivery rate. An
increase of the scale factor reduces the pump speed, and a de-
crease increases the pump speed.
A test infusion line (OIL test infusion line with an Infusomat® fmS
adjusted for the Original Infusomat® line, and a Space calibration
line with an Infusomat® fmS set for the Infusomat® Space Line is
to be used for determining the correction value
.
1.
OK activates the function.
2.
The value can be changed with the entry keyboard
3.
Acknowledge with YES.
4.
Return to initial function with END.
5.
Quit the Service Program and save changes with YES.
6.
Switch on unit and check delivery rate
If necessary repeat the delivery rate measurement.
Summary of Contents for Infusomat fmS
Page 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10...
Page 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes...
Page 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes...
Page 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes...
Page 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes...
Page 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes...
Page 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes...
Page 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes...
Page 84: ...Index 11 11 2 Infusomat fmS 2 3 gb...
Page 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes...