Procedural Instructions on the TSC
8
8 - 8
Infusomat® fmS 2.3 gb
13. Compare value with the threshold value specifications of the
TSC.
Battery Check
1.
Disconnect the device from the mains supply during opera-
tion.
2.
The symbol for the mains plug on the LC display goes off. An
error message is not triggered and operation of the unit is
continued.
3.
Reconnect unit to the mains.
4.
The symbol for the mains plug is displayed on the LC display.
An error message is not triggered and operation of the unit is
continued.
5.
Switch device off.
6.
Disconnect unit from mains.
7.
Switch on unit.
A self-test is carried out.