Service Program
3
Infusomat® fmS 2.3 gb
3 - 5
2.
Return to initial function with END.
Pump Head Cycles
Function 220.0
Display of the pump head cycles (delivered volume).
1.
OK activates the display.
2.
Return to initial function with END.
Operating Alarms
Function 230.0
The last 20 operating alarms can be recalled.
They are displayed as 16 bit binary codes and each bit position can
be set from 0 to 1.
1.
OK activates the alarm display.
2.
Display operating alarms -01 to -20 with the (+) and (-) key.
3.
Delete operating alarms with CLR.
4.
Return to initial function with END.
Unit alarms will also be displayed from software version IFMe,
IFME on
Test
Air Inline Sensor
Function 300.0
After exchange check the function of the air inline sensor.
See TSI-List for permissible check values
1.
Press OK. The received signal amplitude is displayed as a
measured value.
(The test value with NEXT is not important).
2.
Insert an infusion line filled with air and check the maximum
permissible air value.
3.
Insert an infusion line filled with fluid and check the mini-
mum permissible water value.
4.
Return to initial function with END.
Fig.: 3 - 5
0
Battery empty (battery alarm)
1
Pump cover open
2
Drop alarm
3
Air alarm
4 Occlusion
alarm
5
Expired standby time
6
CC alarm (interface)
7
KOR end alarm
8 to 14
free
15
Operating alarm
Bit
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Code
1 0 0 0 0 0 0 0 0 0 0 0 0 0 1 0
Example for alarm ”Pump cover open”
Summary of Contents for Infusomat fmS
Page 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10...
Page 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes...
Page 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes...
Page 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes...
Page 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes...
Page 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes...
Page 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes...
Page 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes...
Page 84: ...Index 11 11 2 Infusomat fmS 2 3 gb...
Page 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes...