Rev. 2.8 Date 2019-02-04
Software version 1.00
Page 8-4
CCS
Chapter 8
System data parameters
Section 1 – Operating Instructions
8.8
System data, menu item 6
Menu item 6 can be seen in all operating modes. It includes the option of changing the system parameters
set by the service staff from B. Braun Avitum AG (B. Braun) during commissioning. Changes may only be
carried out by staff who have been authorized to do so.
8.8.1
Display of system data, menu 6A display
ATTENTION The input of incorrect values may impair the proper functioning of the control or the
operational safety of the system.
Settings must be carried out only by authorised technical staff.
If operating and fault lamp flash alternately, the safety chain has been
interrupted.
The initial menu offers the user the selection of having the set system
data displayed “A Display” or entering these by pressing “B Program”.
6 System data
6.A Display
6.B Program
Esc
Enter
Menu items 6.1 – 6.12
The sub-items 6.1 – 6.2 show the user the current operating values and
set control parameters.
Menu 6.1A to 6.6A
The reference and actual pressures and current pump speed are
displayed as a percentage value of the permissible maximum speed.
System type / software version
Menu item 6.7
verifies the system type.
The system type is indicated during the initial test.
Z
System type ‘H’ or ‘P’.
x
Number of supply groups (number of concentrates to be pumped).
y
Number of supply lines per group (number of floors to be supplied).
The current software version is indicated in the third line.
6.1 Concentrate 1.1
6.2 Concentrate 1.2
6.3 Concentrate 2.1
6.4 Concentrate 2.2
6.5 Concentrate 3.1
6.6 Concentrate 3.2
6.7 System type
6.8 Routine check
6.9 Hours system
6.10 Economy mode
6.11 Remote - on/off
6.12 Control param.
Esc
Enter
6.1 Concentrate 1.1
Ref=X.XX Act=X.XXbar
rpm pump= XXX%
Esc
6.2 Concentrate 1.2
Ref=X.XX Act=X.XXbar
rpm pump= XXX%
Esc