Rev. 2.8 Date 2019-02-04
Software version 1.00
Page 8-6
CCS
Chapter 8
System data parameters
Section 1 – Operating Instructions
8.8.2
System data menu 6B Program
When menu 6.B Program is selected, the following screen appears:
A 4-digit service code must be entered for the system data to be
changed.
Menu 6.1B to 6.6B
Pressure
0.10 to 5.00 bar (D = 0.05 bar).
rpm min
Minimum speed = 0 to 99 % (D = 1 %).
rpm max
Maximum speed = 10 to 100 % (D = 1 %).
Input range 1.1 / 1.2 / 2.1 / 2.2 / 3.1 / 3.2.
Service interval
Input range: 0 to 12 months (0 = -- = off).
System hours
Reset = Reset of system hours to 0.
Economy mode
Each individual valve can go into economy mode to save energy during
the holding phase and to prevent local valve warming.
Economy mode solenoids on/off / input range 0/1
Remote control remote on/off
Input possibilities:
Pulse trigger / slope trigger.
(Positive slope means system on; negative slope = system off.)
Control parameters
To adjust the control parameters of every concentrate line (1.1 – 3.2)
Enter next number
Kp Proportional amplification 0.1 to 1.0 (D = 0.1).
Ki Integration time constant 20 to 980 ms (D = 20).
6 System data
Enter code
XXXX
Esc + - Enter
6.1 Concentrate 1.1
Press.val. = X.XXbar
rpm min=XX max=XXX %
Esc + - Enter
6.8 Routine check
XX months
Esc
6.9 Hours system
XXXXXX hours
Esc Reset
6.10 Economy mode
SV 11 12 21 22 31 32
X X X X X X
Esc
6.11 Remote - on/off
XXXXXXXX
Esc
6.12 Control param.
Line X.X
Kp=X.X Ki=XXXms
Esc