Maintenance and troubleshooting of this instrument
Azbil Corporation
5-2
Model: MVC10A/10F - Multivariable Air Flow Meter
5-1:
Maintenance
Disassembling and assembling
Disassembling
Disassembling of the converter (See Figures 1-4 and 1-5.)
CAUTION
•
When opening or closing the cover for the converter, pay adequate attention not to allow dust and
rainwater to enter the inside of this instrument.
WARNING
• Before opening the cover, not only for disassembling, but also for any other purposes, be sure to
turn off the power. Otherwise an electric shock may result.
•
Whenever the system is energized, never perform wiring work. Otherwise an electric shock may
result.
Removing the cover
The cover is fixed with M4 screws. Loosen them with a Philip's or flat head
screwdriver, and then remove the cover.
Removing the cover for power supply unit
A cable is fed through the gap in the power unit cover and connected to the LCD
board. First remove the connector. The power unit cover is fixed to the case with M3
screws, each with a Philip’s head, at two spots, which should be removed.
Removing the terminal board and main board
The terminal board is fixed to the case with M3 screws, each with a Philip's head, at
four spots.
After removing the terminal board, you will find the main board, which is fixed to the
case with M3 screws, each with a Philip’s head, at two spots. Before removing them,
conduct the static charge removal operation.
Exercise care in this operation not to damage the cable connection of the detector
body and FPC.
CAUTION
• Before and after the removal of the terminal board and main board, be sure to perform the static
charge removal procedure. Otherwise, the performance of the system may be affected.
Assembling the converter
Use the steps for disassembling the converter in the reverse order for the assembling.
When fixing the terminal with screws, pay attention to avoid excessive screw
tightening that may damage them. Also take necessary care not to pinch the cable
connecting the FPC and LCD with screws or the cover.
Summary of Contents for MVC10A
Page 1: ...The Multivariable Air Flow Meter Model MVC10A 10F User s Manual CM2 MVC200 2001 ...
Page 10: ...Table of Contents ...
Page 59: ......
Page 62: ......
Page 64: ......