27
Test Stimulation During Lead Implantation
During tined lead implant procedures, the CP is used to
deliver test stimulation to the foramen needle and the
tined lead. The test stimulation allows confirmation that
the needle and lead are placed close to the sacral nerve.
This section describes how to use the Lead Placement
module of the CP during a tined lead implant procedure.
Press “Lead Placement” (
①
) on the
Home
screen to
1
start foramen needle test stimulation.