Controlling Stimulation Amplitude
23
④
A slider bar indicates the level of stimulation and can
be moved up or down to make large changes to the
stimulation amplitude.
Note:
The slider bar can only
be used to change stimulation amplitude when
stimulation is off.
2
2
⑤
The gray box between the up and down arrows
represents the programmable range of stimulation
amplitude, which is 0 to 12.5mA. At high electrode
3
impedances, the delivered stimulation amplitude may
be lower than the programmed amplitude.
⑥
When stimulation is on, the area under the slider bar
will turn dark gray with horizontal blue bars.
1
5
⑦
Stimulation may be set to increase amplitude
automatically if "Auto" is selected.
Stimulation amplitude can be adjusted in two modes:
manual or automatic.
Manual amplitude adjustment
6
Manual amplitude adjustment is the default mode and is
4
active when "Auto" (
⑦
) is in the light gray, inactive state.
In the manual amplitude adjustment mode, stimulation
8
3
amplitude is changed by moving the slider bar (
④
) or by
pressing the up and down arrows (
③
). Stimulation is
7
7