Controlling Stimulation Amplitude
26
Stimulation output ranges
Maximum stimulation current is dependent on the stimulation settings and impedance values. In the event of high
impedances values or high frequency and pulse width, the output stimulation amplitude may be less than the programmed
stimulation amplitude. Typical settings are 14 Hz and 210 us with impedance from 600 – 1500 Ohms.
Load
(Ohms)
Achievable amplitude
(mA)
14 Hz, 210 µS
(nominal stimulation)
50 Hz, 300 µS
(moderate
stimulation)
130 Hz, 450 µS
(high stimulation)
5000
2.5
2.2
2.0
2000
5.0
4.5
2.5
700
12.5
9.0
3.5
400
12.5
12.5
4.0
Impedances below 400 is considered a short circuit and impedances above 6000 is
considered an open circuit.
If the stimulator is delivering less current than programmed on the CP, when stimulation is active the blue number above the
stimulation bar shows the actual delivered stimulation current.
If stimulation amplitude is being adjusted and no change in patient response is observed, the user should compare the
delivered current (blue display above the stimulation bar when stimulation is on) to the programmed current (gray display
above the stimulation bar when stimulation is off) to determine if the system is delivering the programmed current.