Device Specifications
11
Shipping and Storage environment
The following lists the appropriate temperature, humidity, and pressure condition for shipping and storage of the Axonics
CP:
Temperature (short term: 3 days): -25 °C to 70 °C
Temperature (long term): 20 °C to 30 °C
Humidity (short term: 3 days): 15% to 95%
Humidity (long term): 30% to 85%
Pressure (short term: 3 days): 57 kPa to 106 kPa
Pressure (long term): 70 kPa to 106 kPa
If the CP is exposed to extreme temperatures, it may be permanently damaged and should not be used, even if it
has returned to a temperature that is within the specified operating range.
Maintenance
At least once a year, the CP should be inspected for visible damage and should be charged and powered on to confirm the
Log-In
screen is still accessible. Significant physical damage or an inability to power on the device should be reported to the
manufacturer and the device should not be used.
Handling and disposal
Cleaning:
The CP can be wiped with a cloth lightly moistened with water. No other cleaning agents should be used.
Replacement:
If the CP is lost or not working, contact Axonics.
Disposal:
Do not incinerate the CP as the battery may explode. If the CP is no longer needed, contact Axonics to
return the device.