12
Fig. 5.
Connections light package 2-channels version 1
1
,
2
Connections for light guide
3
,
4
Connections for ATMOS HL 21 LED and
ATMOS LS 21 LED
1 2 3 4
3.0 Setting up and starting up
3.6 Starting up
3.6.1 Connection to electrical power line
• Join the inlet connector for non-heating apparatus (
1
,
fig. 4) with the earth socket outlet using the power cable
supplied.
3.6.2 Waste water connection (with optional
automatic secretion canister evacuation)
• Connect the waste water connector (
2
, fig. 4) to the
water drain (
2
, fig. 3).
3.6.3 Connection of suction hose
• Thread the suction hose through the hole on the side into
the service compartment and insert it onto the secretion
hose connector (
5
, fig. 2).
)
The transport locking screws on the unit's baseplate must
be removed by skilled staff prior to starting up! To do this,
the rear panel must be opened and the two front screws
must be unscrewed. The pump unit is removed from the
rear locking device by pulling it forwards.
3.6.4 Connection of the light guide and light
source
The unit has different connections depending on the
configuration:
Light package LED economy:
• 2 connections for ATMOS HL 21 LED and ATMOS LS 21
LED
Light package 2-channel with 2 or 4 channels:
Version 1 (fig. 5):
• 2 connections for light guide
• 2 connections for ATMOS HL 21 LED and ATMOS LS 21
LED
Version 2:
• 2 or 4 connections for light guide
Version 3:
• 2 or 4 connections for ATMOS HL 21 LED and ATMOS
LS 21 LED
Exclusively connect ATMOS HL 21 LED and ATMOS LS 21
LED to the connections for ATMOS HL 21 LED and ATMOS
LS 21 LED.
Unsuitable application parts may result in an
electric shock or damage.
Cardiac arrhythmia and even
death are possible.
• Connect the light guide or light source.
)
If you require several adapter sleeves (e.g. Olympus,
Pentax, Wolf, etc.), they may be interchanged as
necessary (see chapter 8.2).
)
So that the fibre-optic cables do not rest on the floor, they
may be threaded through behind the handle support.
!