![Atmos ATMOS C 21 Operating Instructions Manual Download Page 4](http://html.mh-extra.com/html/atmos/atmos-c-21/atmos-c-21_operating-instructions-manual_3003587004.webp)
4
1.2 Intended use
1.2.1 ATMOS C 21
Name:
ATMOS C 21
Main functions:
• Suction
• Compressed air system for medication spraying and the
Politzer manoeuvre
• Ear irrigation
• Electric power supply for LED light handles and LED
headlight
• Light sources
• Storage and heating of endoscopes
• Mirror quick heater
• Instrument deposit and heating
Medical indications / application:
Standard ENT examination and/or therapy
Specification of the main function:
• Suction at 40 l/min / -76 kPa
• Alternatively suction at 55 l/min / -98 kPa
• Compressed air for medication spraying, max. 2 bar
• Compressed air for the Politzer manoeuvre, regulated
• Ear irrigation with 37 °C ± 1 °C, max. 500 ml/min
• Electric power supply for LED, 700 mA
• Light source LED for light guide
Application organ:
Mouth to pharynx, auditory canal to the ear drum and the
nasal cavities
Application time:
• ENT unit: Short term use (up to 30 days)
• Suction / Compressed air / ear rinsing / light source:
Temporary application on the patient (less than 60
minutes)
Application site:
Application sites are clinics and practices for ENT doctors
and phoniatrists. The examination and/or therapy with
the ENT unit may only be executed by medically trained
persons.
Contraindications:
May not be used for irrigation of the paranasal sinuses. The
ear irrigation should not be applied to an infected auditory
canal or a perforated eardrum.
The product is:
active
Sterility:
The ENT unit is no sterile product.
Single-use product / reprocessing:
The ENT unit is intended for multiple use. The device and
parts of the accessories are reusable. For information on
reprocessing, cleaning and disinfection, please see the
operating instructions.
1.2.2 ATMOS C 31
Name:
ATMOS C 31
Main functions:
• Suction
• Compressed air system for medication spraying and the
Politzer manoeuvre
• Ear irrigation
• Electric power supply for LED light handles and LED
headlight
• Light sources
• Storage and heating of endoscopes
• Mirror quick heater
• Instrument deposit and heating
Medical indications / application:
Standard ENT examination and/or therapy
Specification of the main function:
• Suction at 40 l/min / -76 kPa
• Alternatively suction at 55 l/min / -98 kPa
• Compressed air for medication spraying, max. 2 bar
• Compressed air for the Politzer manoeuvre, regulated
• Ear irrigation with 37 °C ± 1 °C, max. 500 ml/min
• Electric power supply for LED, 700 mA
• Light source LED for light guide
Application organ:
Mouth to pharynx, auditory canal to the ear drum and the
nasal cavities
Application time:
• ENT unit: Short term use (up to 30 days)
• Suction / Compressed air / ear rinsing / light source:
Temporary application on the patient (less than 60
minutes)
Application site:
Application sites are clinics and practices for ENT doctors
and phoniatrists. The examination and/or therapy with
the ENT unit may only be executed by medically trained
persons.
Contraindications:
May not be used for irrigation of the paranasal sinuses. The
ear irrigation should not be applied to an infected auditory
canal or a perforated eardrum.
The product is:
active
Sterility:
The ENT unit is no sterile product.
Single-use product / reprocessing:
The ENT unit is intended for multiple use. The device and
parts of the accessories are reusable. For information on
reprocessing, cleaning and disinfection, please see the
operating instructions.