7
• The ATMOS C 21 / C 31 may not be operated with units
not complying with the requirements of standard EN
60601-1 "Medical Electrical Equipment" and EN 60601-
1-2 "Electromagnetic compatibility (Medical Electrical
Equipment)".
• The warranty will be rendered invalid in case of
damages caused due to the utilization of accessories or
consumables which are not approved by ATMOS for use
with this unit.
• ATMOS is not liable for personal injury and damage to
property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is
not being observed,
- assembly, new settings, alterations, extensions
and repairs have been carried out by personnel not
authorised by ATMOS.
• Please pay also attention to the safety information in
following chapters.
• Please note:
A medical insulating transformer with earth leakage
monitor or any similar safety system acc. to EN 60 601-1
is required, if several devices are connected over one
common power supply. The transformer must correspond
to the power consumption of all the devices to be
connected.
• When connecting several devices on one grounding
receptacle, the allowed strain and leakage current have
to be observed!
• Never leave the patient unattended at the treatment unit.
•
It is not allowed to use flammable substances with the
device.
2.0 For your safety