L. SOLENOID VALVE ASSIGNMENTS
The cabinet is equipped with three “Solenoid Valves” for automatic operation and
delivery of air, DI water, and salt solution to cabinet components. They are listed as
follows:
Solenoid #
Description
1
DI Water to Bubble Tower
When opened, it will allow DI water into the Bubble Tower.
This solenoid is controlled by the Bubble Tower Level Switch.
2
Air to Bubble Tower (Moist Air )
When opened, it will allow compressed air into the Bubble
Tower. This solenoid is controlled by the items selected by
the operator for a specific test interval.
3
Salt Solution to Fog Nozzle Feed Reservoir
When opened, it will allow Salt Solution to fill the fog nozzle
feed reservoir. This solenoid is controlled by the Fog Nozzle
feed reservoir float switch.
Note: On a BCX 3000 or 4000,
Solenoid #3 is located under the second
solution transfer pump. This solenoid
will be located here.