PRODUCT APPLICATION

C

onstruCtion

•  The water purifier is designed to mount horizontally. 
•  The water purifier’s removable chamber head design facilitates ease of maintenance. A drain port on the chamber aids in 

draining of the purifier.

•  The water purifier’s chamber and chamber head are passivated and electropolished type 316 Stainless Steel.
•  The ballast housing is a combination of Stainless Steel Type 304 and Aluminum Alloy.
•  Coated chambers are available for uses with special applications, consult Factory. Coated chambers are not recommended for 

use in drinking water applications or in applications where the treated product is consumed.

•  The dual action wiper mechanism allows for quick and easy quartz sleeve cleaning, without interrupting service.

•  EASY-OFF

 End Caps enable quick and easy lamp change, without disconnecting from the water supply or draining the 

purifier. No tools are required. 

Always disconnect electrical power when changing lamp.

P

rinCiPle

 

of

 o

Peration

The

 SANITRON

®

 

design has been carefully conceived to provide adequate germicidal dosage throughout the disinfection 

chamber. The dosage, as it applies to ultraviolet disinfection, is a function of time and the intensity of ultraviolet radiation to 

which the water is exposed. The exposure time, in seconds, is the total time it takes the water to flow through the disinfection 

chamber exposing it to the germicidal lamp. Exposure time is related to the flow rate; the higher the flow rate, the lower the 

exposure time or the lower the flow rate, the higher the exposure time. The ultraviolet intensity is the amount of energy, per unit 

time, emitted by the germicidal lamp. The dosage is the product of ultraviolet intensity and the exposure time. The operation of 

the 

SANITRON

®

 is as follows:

•  Water enters the purifier and flows into the annular space between the quartz sleeve and the chamber wall.
•  Suspended microorganisms are exposed to the ultraviolet rays emitted by the germicidal lamp.
•  The translucent sight port, or optional ultraviolet monitor, provides visual indication of germicidal lamp operation.
•  The dual action wiper mechanism facilitates periodic cleaning of the quartz sleeve without disassembly or interruption of 

purifier operation. 

•  Water leaving the purifier is instantly ready for use, no further contact time is required.

l

imitation

 

of

 u

se

The water purifier is intended for the use with visually clear water, not colored, cloudy or turbid. See 

“Water Quality”

 section 

below. The water purifier is 

NOT

 intended for the treatment of water that has an obvious contamination or intentional source, 

such as raw sewage; nor is the purifier intended to convert wastewater to microbiologically safe drinking water.

W

ater

 Q

uality

Water quality plays a major role in the transmission of germicidal ultraviolet rays. It is recommended that the water does not 

exceed the following maximum concentration levels:

Table 1 - Maximum Concentration Levels

Turbidity

< 1 NTU

Manganese

0.05 mg/1

Total Suspended Solids

10 mg /1

pH:

6.5 - 9.5

Color:

None

Hardness

6 GPG or 102.6 PPM

Iron

0.3 mg/1

Tannins:

< 0.1 ppm (0.1 mg/l)

UV Transmission

>85% per cm*

* Contact Factory for recommendations on applications where UV transmission is < 85%

Effectively treating water with higher concentration levels than listed above can be accomplished, but may require added 

measures to improve water quality to treatable levels. If, for any reason, it is believed the ultraviolet transmission is not 

satisfactory, contact the factory.

Summary of Contents for Sanitron S17A

Page 1: ...d Follow All Safety Instructions Save These Instructions Models S17A S23A S37C S50C Extensive Product Information Available at Ultraviolet com BuyUltraviolet com Models S37C and S50C Installation Oper...

Page 2: ...ns when packaging equipment to prevent damage Care fully inspect and report all damages Do not install damaged equipment Follow all instructions on any labels or tags Carefully inspect all packing mat...

Page 3: ...out proper electrical ground 12 Do not exceed water purifier s maximum rated flow capacity 13 Do not exceed maximum operating pressure of 100 PSI 14 Read and follow all notices and warnings on the wat...

Page 4: ...cidal lamp The dosage is the product of ultraviolet intensity and the exposure time The operation of the SANITRON is as follows Water enters the purifier and flows into the annular space between the q...

Page 5: ...a where any possible condensation or leakage from the water purifier any purifier accessory and or plumbing will not result in damage to the area surrounding the water purifier For added protection it...

Page 6: ...or with the optional wall mounting kit The purifier should not be solely supported by its plumbing connections 4 The water purifier must be connected to the cold water line only Inlet water temperatu...

Page 7: ...r Figure 4 NOTE Be sure to keep the large O Ring on the chamber head when removing this from the chamber 4 Remove PVC pipe and discard REMINDER be sure that the Teflon washer and O Ring from both ends...

Page 8: ...8 Figure 3 Initial Installation of Quartz Sleeve Figure 4 Initial Installation of Quartz Sleeve TELFON WIPER SEGMENTS GLAND FITTING QUARTZ SLEEVE WIPER ROD ASSEMBLY CHAMBER HEAD LARGE RUBBER O RING...

Page 9: ...in conjunction with the Ultraviolet Monitor Safety Sensor or Time Delay Mechanism this valve prevents water flow through the water purifier when an abnormal condition is detected or in the event of p...

Page 10: ...culty due to deposits on the quartz sleeve your dealer will be able to recommend suitable pretreatment to reduce or eliminate the offending contaminant If quartz sleeve is clean clear then frequency o...

Page 11: ...lamp straight out without angling until completely clear of quartz sleeve Lamp and quartz sleeve are easily damaged Exercise care when handling 7 Reinstall lamp in reverse order Germicidal ultraviolet...

Page 12: ...is easily accomplished using the dual action wiper mechanism 1 Lift wiper clip up and away from wiper rod 2 Grasp wiper knob and gently pull out away from the purifier until it reaches its stop 3 Push...

Page 13: ...static gland nut If so remove Teflon washer from static gland nut before proceeding 6 Carefully remove quartz sleeve from chamber NOTE It is advisable to support the quartz sleeve on the opposite end...

Page 14: ...ep ballast housing and mounting hardware in a clean dry area 2 Carefully remove as much of the broken quartz sleeve as possible from each end of the chamber 3 Remove chamber head clamp by using a 7 16...

Page 15: ...g located in the center of the disinfection chamber and protrudes into the chamber 1 Disconnect power to water purifier 2 Shut off water supply to the water purifier via the inlet and outlet shut off...

Page 16: ...tenance section Quartz sleeve sealing O Ring s worn damaged Shut down purifier drain and remove static gland nut replace sealing O Ring See Quartz Sleeve Installation or Replacement in the Maintenance...

Page 17: ...dio Alarm Optional Optional Optional Optional Solenoid Valve Optional Optional Optional Optional Time Delay Mechanism Optional Optional Optional Optional Elapsed Time Indicator Optional Optional Optio...

Page 18: ...18 REPLACEMENT PARTS SANITRON Models S17A S23A Figure 11 Exploded View S17A S23A...

Page 19: ...15 Chamber 1 25 1524D1 1 25 1523E1 16 Wiper Knob 1 25 1222 1 25 1222 17 Wiper Lock Spacer 1 25 1512A1 1 25 1512A1 18 Wiper Clip 1 25 1507C1 1 25 1507C1 19 Dynamic Gland Wiper Lock 1 25 1510C1 1 25 151...

Page 20: ...20 SANITRON Models S37C S50C Figure 12 Exploded View S37C S50C...

Page 21: ...acer 1 25 1512A1 1 25 1512A1 18 Wiper Clip 1 25 1507C1 1 25 1507C1 19 Dynamic Gland Wiper Lock 1 25 1510C1 1 25 1510C1 20 Drain Plug 1 27 1216 1 27 1216 21 Screw 1 4 20 x 3 8 long 4 50 1034 4 50 1034...

Page 22: ...22 MAINTENANCE NOTES...

Page 23: ...23 MAINTENANCE NOTES...

Page 24: ...lamps and Other parts One 1 year Limited Warranty from the date of original purchase Within the warranty period we shall repair or replace such products which are returned to us with shipping charges...

Reviews: