11
IN ORDER TO PERFORM THIS TASK, BE SURE TO WEAR THE FOLLOWING SAFETY
EQUIPMENT: SAFETY GLASSES OR A FACE SHIELD, AS WELL AS GLOVES.
1.
Disconnect power to water purifier.
2. Remove both
EASY-OFF
™
End Caps by pulling each cap off static gland nut. Slide each end cap along the wire away from
the socket.
3. Carefully withdraw lamp approximately 2 inches from chamber while feeding lamp socket and lead wire on opposite
end of chamber.
4. While holding lamp end, carefully remove lamp socket on end now exposed.
5. Next, carefully slide lamp back into chamber, until approximately 2 inches of the lamp is exposed on the opposite end. Hold
lamp and remove lamp socket.
6. Lamp should now be disconnected on both ends. Carefully remove lamp from chamber. Be sure to withdraw lamp straight
out without angling until completely clear of quartz sleeve.
Lamp and quartz sleeve are easily damaged. Exercise care when handling.
7. Reinstall lamp in reverse order.
Germicidal ultraviolet rays are harmful to eyes and skin. Do not restore power to water purifier until
lamp and both
EASY-OFF
™
End Caps have been properly reinstalled.
D
isPosal
of
m
erCury
a
DDeD
l
amPs
Germicidal ultraviolet lamps, like standard fluorescent lamps contain small amounts of mercury. Mercury added
lamps should not be placed in the trash. Dispose of properly. For further information regarding the disposal and
recycling of lamps containing mercury, along with Federal and State requirements visit LampRecycle.org. For more
information on
STER-L-RAY
®
Germicidal Ultraviolet Lamps, visit Ultraviolet.com or BuyUltraviolet.com.
l
amP
i
nstallation
or
r
ePlaCement
Step 2 - Remove End -Cap
Step 6 - Remove Lamp
Step 4 & 5 - Remove Sockets
Step 3 - Withdraw Lamp
Figure 5 - Lamp Replacement