Atlantic Ultraviolet Sanitron S17A Owner'S Manual Download Page 15

15

10.  Replace the head clamp around the flared end of the head and chamber. Install the 1/4” nut and tighten, using a 7/16” wrench, 

until approximately 3/4” to 7/8” of the bolt protrudes past the nut.

11.  Center the quartz sleeve in the chamber, making sure the quartz sleeve protrudes an equal distance past each threaded gland 

fitting, of the chamber. 

12.  Re-install O-Rings, Teflon

®

 washers, and static gland nuts. Be sure O-Rings are placed on quartz sleeve before Teflon

®

 

washer. Tighten static gland nuts firmly by hand only, 

DO NOT USE HAND TOOLS

. Tightening with hand tools is likely to 

cause quartz sleeve to break. 

13.  Re-assemble ballast housing to purifier chamber, using hardware removed in Step 1. 

NOTE:

 When re-assembling ballast 

housing to purifier chamber, electrical power cord should exit ballast housing from the end mounted opposite the drain plug.

14.  When all connections are complete, allow water to enter the water purifier at a low flow rate until the purifier is pressurized. 

With the purifier pressurized, it should be checked for leaks.

15.  See “

Lamp Installation or Replacement

” section on 

page 11

 to properly re-install the lamp into the water purifie

r. 

o

Ptional

 g

uarDian

tm

 u

ltraviolet

 m

onitor

 s

ensor

 P

roBe

 C

leaning

Figure 9 -  Ultraviolet Monitor and Sensor Probe

Figure 10 -  Sensor Probe

If after cleaning the quartz sleeve, there is no significant improvement in the ultraviolet intensity, as shown on the intensity meter, 

it may be necessary to clean the Ultraviolet Monitor’s sensor probe. The sensor probe body mounts in a fitting, located in the 

center of the disinfection chamber, and protrudes into the chamber.
1. 

 Disconnect power to water purifier.

2.  Shut off water supply to the water purifier via the inlet and outlet shut off valves. Remove drain plug and drain the chamber. 

Once the chamber is drained, remove any old sealing tape from the threads of the drain plug, rewrap with 1/2” wide Teflon

®

 

thread sealing tape, reinstall and tighten the drain plug.

3.  Disconnect power to the Ultraviolet Monitor; remove from chamber by loosening the two set screws on the aluminum collar 

and lifting it free from probe body.

4.  Unscrew the sensor probe and remove from the chamber.
5.  Remove the quartz rod sensor probe cap, from the probe body. Take care not to damage the quartz rod, O-Ring or the threads 

of the probe body. 

NOTE:

 It is recommended, when servicing the sensor probe, to work in a clean, dry area.

6.  Once the quartz rod is removed, clean with alcohol or a mild detergent, and rinse with clean water. Stubborn stains usually 

can be removed with a dilute hydrochloric acid. 

NOTE:

 Follow all manufacturer’s instructions and precautions when 

handling chemicals. Once the quartz rod has been cleaned, handle the rod by the sides, to avoid getting fingerprints on the 

quartz rod faces.

7.  Clean the probe body, by removing any dirt or deposits on all surfaces. O-Rings should be inspected and can be replaced if 

worn or damaged.

8.  Reassemble, replacing the O-Rings, quartz rod and securing in place with the quartz rod sensor probe cap. Tighten the quartz 

rod sensor probe cap by hand only, 

DO NOT USE HAND TOOLS

. Tightening with hand tools may damage the quartz rod 

or O-Ring seal.

9.  Reinstall sensor probe into the center fitting of the chamber and hand tighten. 

(Continued on Page 16)

Summary of Contents for Sanitron S17A

Page 1: ...d Follow All Safety Instructions Save These Instructions Models S17A S23A S37C S50C Extensive Product Information Available at Ultraviolet com BuyUltraviolet com Models S37C and S50C Installation Oper...

Page 2: ...ns when packaging equipment to prevent damage Care fully inspect and report all damages Do not install damaged equipment Follow all instructions on any labels or tags Carefully inspect all packing mat...

Page 3: ...out proper electrical ground 12 Do not exceed water purifier s maximum rated flow capacity 13 Do not exceed maximum operating pressure of 100 PSI 14 Read and follow all notices and warnings on the wat...

Page 4: ...cidal lamp The dosage is the product of ultraviolet intensity and the exposure time The operation of the SANITRON is as follows Water enters the purifier and flows into the annular space between the q...

Page 5: ...a where any possible condensation or leakage from the water purifier any purifier accessory and or plumbing will not result in damage to the area surrounding the water purifier For added protection it...

Page 6: ...or with the optional wall mounting kit The purifier should not be solely supported by its plumbing connections 4 The water purifier must be connected to the cold water line only Inlet water temperatu...

Page 7: ...r Figure 4 NOTE Be sure to keep the large O Ring on the chamber head when removing this from the chamber 4 Remove PVC pipe and discard REMINDER be sure that the Teflon washer and O Ring from both ends...

Page 8: ...8 Figure 3 Initial Installation of Quartz Sleeve Figure 4 Initial Installation of Quartz Sleeve TELFON WIPER SEGMENTS GLAND FITTING QUARTZ SLEEVE WIPER ROD ASSEMBLY CHAMBER HEAD LARGE RUBBER O RING...

Page 9: ...in conjunction with the Ultraviolet Monitor Safety Sensor or Time Delay Mechanism this valve prevents water flow through the water purifier when an abnormal condition is detected or in the event of p...

Page 10: ...culty due to deposits on the quartz sleeve your dealer will be able to recommend suitable pretreatment to reduce or eliminate the offending contaminant If quartz sleeve is clean clear then frequency o...

Page 11: ...lamp straight out without angling until completely clear of quartz sleeve Lamp and quartz sleeve are easily damaged Exercise care when handling 7 Reinstall lamp in reverse order Germicidal ultraviolet...

Page 12: ...is easily accomplished using the dual action wiper mechanism 1 Lift wiper clip up and away from wiper rod 2 Grasp wiper knob and gently pull out away from the purifier until it reaches its stop 3 Push...

Page 13: ...static gland nut If so remove Teflon washer from static gland nut before proceeding 6 Carefully remove quartz sleeve from chamber NOTE It is advisable to support the quartz sleeve on the opposite end...

Page 14: ...ep ballast housing and mounting hardware in a clean dry area 2 Carefully remove as much of the broken quartz sleeve as possible from each end of the chamber 3 Remove chamber head clamp by using a 7 16...

Page 15: ...g located in the center of the disinfection chamber and protrudes into the chamber 1 Disconnect power to water purifier 2 Shut off water supply to the water purifier via the inlet and outlet shut off...

Page 16: ...tenance section Quartz sleeve sealing O Ring s worn damaged Shut down purifier drain and remove static gland nut replace sealing O Ring See Quartz Sleeve Installation or Replacement in the Maintenance...

Page 17: ...dio Alarm Optional Optional Optional Optional Solenoid Valve Optional Optional Optional Optional Time Delay Mechanism Optional Optional Optional Optional Elapsed Time Indicator Optional Optional Optio...

Page 18: ...18 REPLACEMENT PARTS SANITRON Models S17A S23A Figure 11 Exploded View S17A S23A...

Page 19: ...15 Chamber 1 25 1524D1 1 25 1523E1 16 Wiper Knob 1 25 1222 1 25 1222 17 Wiper Lock Spacer 1 25 1512A1 1 25 1512A1 18 Wiper Clip 1 25 1507C1 1 25 1507C1 19 Dynamic Gland Wiper Lock 1 25 1510C1 1 25 151...

Page 20: ...20 SANITRON Models S37C S50C Figure 12 Exploded View S37C S50C...

Page 21: ...acer 1 25 1512A1 1 25 1512A1 18 Wiper Clip 1 25 1507C1 1 25 1507C1 19 Dynamic Gland Wiper Lock 1 25 1510C1 1 25 1510C1 20 Drain Plug 1 27 1216 1 27 1216 21 Screw 1 4 20 x 3 8 long 4 50 1034 4 50 1034...

Page 22: ...22 MAINTENANCE NOTES...

Page 23: ...23 MAINTENANCE NOTES...

Page 24: ...lamps and Other parts One 1 year Limited Warranty from the date of original purchase Within the warranty period we shall repair or replace such products which are returned to us with shipping charges...

Reviews: