15
Be careful! Damage to the machine!
Origin shall not exceed one atmosphere pressure, otherwise it will damage the machine.
Never use flammable or other mixed gas containing oxygen, otherwise there will
be an explosion hazard. Missed this provision will lead to an accident or damage,
the manufacturer will not bear any responsibility.
Air tank must be properly closed. When not adopting inflated function or machine
is not used, you must turn off the main switch of tank.
Pressure valve on the gas cylinder must set the value of not more than
atmospheric pressure (0.1Mpa), otherwise it will damage the machine.
Inflatable tubular threaded joint diameter machine is 8mm. Connector is located
on the side or back of the machine.
For cylinders used in case of doubt, please contact the manufacturer or supplier.
Note: If you make too many pneumatic vacuum of 0.06Mpa above, began sealing
function, the sealing function may not be completed properly.
6).Connect the vacuum system (for external vacuum pump set)
If you buy inside pump vacuum machine, you do not need to proceed with the
operation.
If you purchased an external vacuum pump type machine, connect proper pump
before using.
5. Start-up & Debugger
1). Control panel
Button panel presentation
DZ series panel (without gas filled function)
No.
Name
Description
A
Power indicator
shows the power status of the machine, it will be light