© Associated Research 2017
41
Connect the black return test lead to the black Return terminal located on the front panel of the instrument.
Connect the sense lead to the Sense(-) terminal located just above the Return terminal.
Clip the end of the black return test lead to the chassis ground of the DUT.
Connect the red high current test lead to the red Current terminal on the front panel of the instrument.
Connect the red high current sense lead to the red Sense(+) terminal.
Connect the end of the red high current test lead clip to the ground/earth contact of the DUT.
Instrument Connections
You may connect the test leads to the receptacles located on the front or back of HYAMP. These receptacles are wired in parallel
and you can use either depending upon the application.
Test Lead Connections
Class I Product
RETURN
CURRENT
RESET
TEST
HYAMP® Series
GROUND BOND TESTER
SENSE +
TEST
SENSE -
My Menu / Select
POWER
DATA
BARCODE