© Associated Research 2017
36
RETURN
CURRENT
RESET
TEST
HYAMP® Series
GROUND BOND TESTER
SENSE +
TEST
SENSE -
My Menu / Select
POWER
DATA
BARCODE
Setup Ground Bond Test Continued
Offset Connections
To perform mOhms or Voltage Auto
Offset short Red Current/Sense+ Lead
with Black Return/Sense- Lead.
Push TEST button. The instrument will
Auto Offset measurements and save the
value to the test step.*
* Note: Perorming either mOhms or Voltage Auto
Offset will set two values, mOhms and Voltage
offset parameters of the test step.
Enter
Offset
value to manually set
Offset or Press TEST to perform auto
offset.
Last option is defaults which will load
GND parameters.
Connect the Red and Black
alligator clips together.
Select test
Frequency
.
Enter text for
Prompt
.