4.11 Error numbers
If an operation or device error occurs the NMG2 will show an error number in the lower display [18].
The error number can be deleted by pressing the button
ADJUST/TEST RUN
[1] or switching off the device.
Meaning of the error numbers
Error number
Description
11
Line interruption:
The NMG2 has detected a line interruption.
12
Printer error:
The built-in printer doesn
’t respond.
51
Measurement aborted:
After the start of the measurement the measurement has been aborted by the
user.
52
Measuring time too long:
No standstill within the max. measurable stop time (5 s).
53
Measurement velocity:
The permissible measurement velocity was exceeded.
54
Measurement acceleration:
The permissible measurement acceleration was exceeded.
91
RAM error:
A defective memory cell has been found.
92
Relay feedback:
The relay is not in the defined switching state.
Relay is defective.
93
Relay time:
The max. permissible relay switching time was exceeded.
Relay is defective.
94
Actuator not connected (actuator button pushed)
Actuator is defective.
95
Actuator slow
Switching is time too long.