6
•
Before entering a hazardous area, close both the battery
door and USB/SD card door, and secure both with lock
screws.
•
The battery access cover is not to be removed while in a
hazardous area.
•
The USB/SD access door is not to be removed while in a
hazardous area.
•
Replace batteries only in a non-hazardous area.
•
Use only Duracell Model MN1500 AA alkaline primary cell
batteries.
•
The USB connection-data transmission and power input- are
restricted to be used in a non-hazardous area only.
•
The non-hazardous area USB apparatus –computer port,
wall supply, etc., connected to the USB port of the HHC,
must be assessed and conform to section 6.2.5 of EN60079-
11. This precaution is to protect the integrity of the safety
components within the HHC, which insures its intrinsic safety
rating while in the hazardous location.
•
When using non- assessed USB equipment, accessory
part number 101C225-01, USB protection device, is to be
inserted between the HHC USB port and the non-assessed
apparatus.
•
Do not power the calibrator by USB cable in a hazardous
area; this action will invalidate the calibrator’s intrinsic safety.
•
Intrinsically safe installation diagrams, entity parameters and
warnings are included in drawing 825A028 included in the
supplemental I.S. manual.
ADDITIONAL WARNINGS CONT.
250 EAST MAIN ST.
STRATFORD, CT 06614
www.ashcroft.com
INTRINSICALLY SAFE/SÈCURITÈ INTRINSÈQUE Exia
CL I, DIV 1, GR A, B, C, D. T4 -20°C
≤
Ta
≤
+50°C
II 1 G FM12ATEX0035X
Ex ia II C T4 Ga -20°C
≤
Ta
≤
+50°C
ENTITY PARAMETERS
Ui=33Vdc Ii=300mA
Ci=0
Li=0 Pi=1.5W
Uo=5.735Vdc Io=586
A Co=46
f L0=1H Po=840
W
See Dwg #825A028 for installation, batteries
and warnings
HAND HELD CALIBRATOR 6D49 E171189
Unit contains a lithium battery. Panasonic #BR1225
is the approved replacement battery. Use of another
battery may present a risk of fire or explosion. See
owners manual
for safety instructions.
WARNING:
0518
Label Part# 238A746-01Rev D
Summary of Contents for ATE-2
Page 1: ...i I M002 10233 ASHCROFT ATE 2 Operation and Maintenance Manual 0 0 5 4 P S I 0 8 6 i n H 2 0...
Page 2: ...ii...
Page 50: ...48 SECTION 15 l FIRMWARE UPDATE CONT...
Page 51: ...49 SECTION 15 l FIRMWARE UPDATE CONT...
Page 55: ...53 SECTION 15 l FIRMWARE UPDATE CONT Note The File Transfer can take up to 15 Minutes...