IMPORTANT NOTE for INTRINSICALLY SAFE APPROVED
VERSIONS OF THE Hand Held Calibrator
1. When access port is open, the device is not operating in
an Intrinsically Safe mode. As such it is not possible to
use the HHC in a hazardous area whenever the USB port
is exposed or if the port access door is not secured with
Philip’s head screw.
2. When using an Intrinsically Safe approved HHC in a
non-hazardous environment, the USB protection adapter
must be used to connect the HHC to a PC’s USB port.
Failure to install the USB protection adaptor can result in
damage to the electronics which could render the HHC no
longer intrinsically safe.
Driver Installation
If the PC being used together with the HHC does not already
have the proper drivers installed it will be necessary to install a
driver to enable communication.
The HHC need not be powered on for the driver installation.
Place the memory stick that came with the HHC in one of the
PCs USB ports.
SECTION 15 m. i.
USB Cable Installation Cont.
SECTION 15 m. ii.
Driver Installation