Electromagnetic Emissions
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 42 of 43
Table 18
Guidance and Manufacturer's Statement - Electromagnetic Immunity
The AR-6480 DualWave Arthroscopy Pump is intended for use in the electromagnetic environment specified below. The
customer or the user of the AR-6480 DualWave Arthroscopy Pump should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
Portable and mobile RF communications
equipment should be used no closer to any part of
the AR-8305 Synergy Resection System, including
cables, than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1.17P
1/2
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
80 MHz to 2.5 GHz
d = 1.17P
1/2
80 MHz to 800 MHz
d = 2.33P
1/2
800 MHz to 2.5 GHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Interference may occur in the vicinity of
equipment marked with the following symbol: