Operation
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 29 of 43
pressure increase of 30% for 60 seconds or to the user defined
parameters.
2.
The RINSE mode will stop when the countdown reaches zero, or if
the user presses the RINSE button a second time.
5.6
Alternative Suction Pathway (ASP) Mode
The AR-6480 pump has an ASP function for minimizing clogging issues. The
ASP MODE is only accessible in INFLOW/OUTFLOW MODE. In ASP Mode
changes the suction pathway from the shaver suction pathway to the cannula
suction pathway at the shaver suction level.
1.
Turn ON ASP MODE in OUTFLOW SHAVER default menu.
2.
Press the RINSE button on the Remote control or Foot Pedal [iv] while
the shaver is activated to enable this function. The Shaver button will
turn Red. The suction pathway will change from the shaver suction
pathway to the cannula suction pathway.
3.
The ASP mode will stop when the shaver is deactivated or by exiting
ASP mode when the user presses the RINSE button a second time on
the Remote Control or Foot Pedal [IV] while the shaver is activated.