Setup
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 19 of 43
4.3
Replacing the Fuses
Replace the 250v 6.3A mains fuse as follows:
1. Disconnect the device from the AC mains.
2. Open the fuse tray in the AC inlet, by pinching the tabs and pulling
outwards.
3. Replace the fuses with 250V 6.3A Line Fuses as noted on the rear panel.
4. Push the fuse holder back into the AC inlet.
5. Ensure that fuse holder is fully seated and that the tabs snap back.
Always use fuses with the correct values to avoid allowing
overcurrent to enter the system.
NOTE: The AR-6480 console incorporates a universal AC input power supply.
A voltage selection switch is not required.
4.4
Electromagnetic Compatibility
This device has passed testing for EMI / RFI radiation and
susceptibility and EMC compatibility. This device may cause
interference to other devices in the near vicinity, if not set up and
used as Arthrex instructs.
To determine if the AR-6480 is causing interference to other devices, power OFF
the AC mains power switch [3] and then ON again.
Try to correct the interference by following one or more of these measures:
1.
Reorient or relocate the receiving device.
2.
Increase the separation between devices.
3.
Connect the device to an outlet on a different circuit than the other
device(s) are connected.
4.
Consult the manufacturer or field service technician for the receiving
device for guidance.
4.5
Basic Setup Procedure for the AR-6480
W A R N I N G !
To assure that correct pressure monitoring occurs, the pump and operative site
MUST be in the same horizontal plane.
NOTE: Section 5.0, Operation, explains how to use the pump.
1.
Place the AR-6480 on a flat, dry surface in the same horizontal plane
as the operative site, such as the AR-6481 Arthrex Arthroscopy pump
cart.