Operation
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 28 of 43
5.3
How to Operate the AR-6480 in INFLOW/OUTFLOW Mode
W A R N I N G !
The Extension or Patient Tubing must be replaced before each new surgical
procedure.
1.
Follow steps as outlined in 5.2.
2.
Add Outflow tube set to AR-6480.
3.
The sensor will detect that the outflow tubing has been added and
change the controls and icons on the TPVD to signify
INFLOW/OUTFLOW mode.
4.
The waste tubing end should be attached to a waste bag or suction
device/container.
5.
The end of the tubing labeled SHAVER should be attached to the
shaver suction port.
6.
The end of the tubing labeled CANNULA can be attached to an
outflow cannula, working cannula, or left unutilized.
NOTE: If the CANNULA tubing end is not used, leave unclamped.
5.4
How to Operate the AR-6480 in LAVAGE Mode
The AR-6480 pump has a LAVAGE function for haemostatic purposes. The
LAVAGE mode is accessible in INFLOW ONLY and INFLOW/OUTFLOW
MODE.
W A R N I N G !
User programmed “Pressure Set” values are increased by as much as 50% to a
maximum of 120 mmHg during the LAVAGE function. Exercise caution to avoid
injury to the patient.
1.
Press the LAVAGE button [Table 3 or I] to enable this function. The
LAVAGE button should turn color and begin a countdown. The
pressure will be increased to the factory default of a 50% increase for
120 seconds or to the user defined parameters.
2.
The LAVAGE mode will stop when the countdown reaches zero, or if
the user presses the LAVAGE button a second time.
5.5
How to Operate the AR-6480 in RINSE Mode
The AR-6480 pump has a RINSE function for irrigation purposes. The RINSE
MODE is only accessible in INFLOW/OUTFLOW MODE.
1.
Press the RINSE button [Table 3 or IV] to enable this function. The
RINSE button should turn color and begin a countdown. The Outflow
will be increased to the factory default of a 300mL/min with a