Setup
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 25 of 43
Roller Housing
4.12.3
The pump motor automatically deactivates when the roller housing door is
opened. A locking mechanism prevents access to the rotating parts while the
device is operating.
Tubing Sensor Coupler
4.12.4
The pump motor automatically deactivates when the tubing is disconnected
from the pump. If the tubing is disconnected during a surgical procedure it must
be replaced by new tubing. Do not reconnect the tubing to the pump as it could
lead to unreliable pressure measurements.
W A R N I N G !
If the tubing is disconnected from the pump in the middle of a procedure it must
be replaced. Do not attempt to reconnect the tubing to the pump as it could lead
to unreliable pressure measurements.