Setup
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 23 of 43
4.
Guide the tubing boot [d] over the rollers and insert the output side of
the tubing boot into the Tubing OUT Guide.
NOTE: The Pump Tubing is connected properly when the green connector [c]
on the Main Pump Tubing is aligned with the green dot [1] on the front
panel of the console.
5.
Close the inflow door.
NOTE: The inflow door locking device must be secure. If the door is not closed
securely an internal safety switch prevents the AR-6480 from operating.
6.
Puncture the fluid bags with the spikes on the tubing. If only one fluid
bag is being used, seal the second fluid line by closing the clamp
nearest the unused spike.
4.8
How to Set Up the Two-Piece Tubing System
NOTE: These instructions describe the procedure to set up the
Extension
Tubing or ReDeuce™ Patient Tubing
.
W A R N I N G !
The Extension or Patient Tubing must be changed for each patient.
1.
The surgical staff removes the sterile Extension or Patient Tubing
from its sterile pack and hands the connector [h or k] for the Pump
Tubing set to the circulating nurse.
2.
The circulating nurse connects the two tubing systems together [h to h
in Figure 5 or k to k in Figure 6].
3.
Attach the sterile connector cap (supplied with each Extension or
Patient Tubing set) to the patient-end of the Pump Tubing
.
NOTE:
Following each surgery, detach and discard the Extension or
Patient Tubing Set
.
W A R N I N G !
The sterile connector cap must be used to cover the Pump Tubing Set connector
after each surgical procedure. This maintains sterility of the Pump Tubing and
assures its safe operation throughout the entire surgical day.
4.9
How to Set Up the Outflow Tubing
1.
Open outflow door completely.
2.
Place looped tubing around the roller assembly.
3.
Pull Outflow Fixture [E] down until it slides into the outflow tubing
receiver.