Care of your MAXI MOVE
38
001
.2
506
0.E
N
r
e
v.
1
1
The frequency of the following actions depends
on how often the product is used.
Unless otherwise stated, follow the cleaning, care
and inspection procedures described in this
section before each and every use.
Sling Cleaning and Care
The slings should be checked before and after
each patient use and if necessary, washed in
strict accordance to the instructions on the sling.
This is especially important when using the same
product for another patient, as it can minimize the
risk of cross-infection. Also refer to sling
instruction sheet MAX81785M-INT.
Slings should not be classified as linen, but as an
accessory to a patient transfer lift and, therefore,
are classified as a medical device accessory.
Mechanical pressure, such as rolling or pressing,
should be avoided during the washing and drying
procedure, as this can damage parts that are vital
to the safe and comfortable operation of the sling.
The strap stretcher cross straps and suspension
straps should be checked and washed if
necessary. Washing and drying temperatures
must not exceed 80°C (176°F). Wash using
normal detergents. Do not iron. Also refer to the
Sling Instruction sheet MAX81785M-INT.
It is essential that the slings, sling loops, straps
and attachment clips are carefully inspected
before each and every use. If the slings, loops or
straps are frayed or the clips damaged, the sling
must not be used and should be replaced
immediately.
Lift Cleaning and Care
Removing visible residues:
1) Remove the battery pack and general parts
such as slings, cushions and suspension
straps (if applicable).
2) Remove visible residues from the MAXI MOVE
using a cloth soaked in water. Start from top
and move downwards.
3) Remove visible residues from the removed
parts and battery pack with a cloth soaked in
water mixed with cleaning fluid.
Cleaning:
1) Using a brush or a cloth soaked in cleaning
fluid, wipe the lift, battery pack and removed
parts vigorously (to remove any deposits).
2) Use a new damped cloth with clean water to
wipe off all traces of cleaning fluid on the lift,
the battery pack and removed parts.
3) If cleaning fluid cannot be removed on some
hard-to-access parts, spray water on the
affected part and wipe off with disposable
towels/cloth. Repeat until all of the cleaning
fluid has been removed.
4) Repeat steps 1 to 3.
5) Let the parts dry.
6) Reinstall the removed parts and battery pack
on the lift.
NOTE: The lift and the sling should be cleaned
between use of different patients and/or when
suspected to be contaminated
.
WARNING: To avoid eye and skin injury never
disinfect the lift or accessories in the presence of
a patient and always use protective gloves. If
contact occur rinse with plenty of water. If eyes
or skin become irritated seek for medical
attention. Always read the material safety data
sheet related to the disinfectant.
NOTE: Pay special attention to areas pointed
below. These are most likely to enclose germs.
Use a smaller brush and/or cotton swab to reach
them.
Care of your MAXI MOVE
MAXI MOVE’s Special Areas to Clean
Fig. 76
Spreader bar
Locking clip
assembly
Handset handle
Control box
membrane
edge