Scale
32
001
.2
506
0.E
N
r
e
v.
1
1
Fig. 62
4) Lift the patient until obstructions are cleared,
such as the bed, chairs, the floor, etc. Allow the
weight reading to stabilize.
Do not press the button again; the number
displayed will be the patient’s weight (see
Fig. 63).
Fig. 63
5) Press the save button if the net weight needs
to be kept in memory. Horizontal bars will
appear above and below the digits on the
screen to show that the save function is
activated. “Save” will stay active until the save
button is pressed again (see Fig. 64).
Fig. 64
Method B - Weighing with the
Patient Already Suspended in the
Sling
If the patient is already on the lift, and a weight
measurement is needed, ensure that the patient
is suspended free and clear of any obstructions
such as the bed, chairs, the floor, etc.
1) Press the scale button to obtain a zero reading
on the display (see Fig. 65).
Fig. 65
2) Complete the transfer of the patient and
remove the patient from the lift. The scale will
display a negative number (see Fig. 66).
Fig. 66
3) Reinstall the sling back on the MAXI MOVE.
Ignore the minus sign preceding the digits on
the screen. Allow the weight reading to
stabilize. The weight shown is the patient’s
actual weight (see Fig. 67).
Fig. 67
4) Press the save button if the net weight needs
to be kept in memory. Horizontal bars will
appear above and below the digits on the
screen to show that the save function is
activated. “Save” will stay active until the save
button is pressed again
.
Save
button
Horizontal bars
show that “Save”
mode is on
CAUTION: If the unit is reset while the patient is
still suspended in the sling, the scale will move
out of its zero range and display “8888.8” to
indicate an error status. Remove the patient from
the MAXI MOVE and reset the unit.
Scale
button