Technical Specifications
42
001
.2
506
0.E
N
r
e
v.
1
1
PRODUCT INFORMATION
MAXI MOVE
Total weight (standard jib)
63.5 kg (140 lb)
Lifting capacity (SWL)
Regular jib: 227 kg (500 lb)
Extended jib: 130 kg (287 lb)
Battery pack weight
5.0 kg (11 lb)
Turning Diameter
1222 mm (48 in)
Minimum door requirement
717 mm (28.25 in) for Standard Base (KMCS
**
) & Low Height Base (KMCL
**
)
770 mm (30 3/8) in) for Extra Low Height Base (KMCE
**
)
Operating forces of controls
2.5- 3N
ELECTRICAL
Degree of protection
IPx7 - Hand control
IP24 - MAXI MOVE
Internally powered
24 Vdc
Battery Type
Rechargeable (Sealed Lead Acid)
Battery Capacity
24 Vdc, 4.0 Ah to 5.5 Ah
Battery charger input
(part #NDA8200)
100 to 240 Vac / 50 to 60 Hz
Up and down current limiting
12 ± 1 Amp.
Duty cycle
15% - max. 2 min. continuous use
Sound power level up
61.9 dBA
Sound power level down
61.7 dBA
Medical equipment
Type BF protection against electrical shock in accordance with IEC 60601-1
ArjoHuntleigh products meet the requirements of Electromagnetic Compatibility
(EMC) as stated in clause 12.5 of the Medical Devices Directive 93/42/EEC.
The MAXI MOVE conforms to UL 60601-1, CSA C22.2 No. 60601.1 - M90 and ISO 10535:2006 (except when
being used in conjunction with a stretcher; see “Using your MAXI MOVE” section on page 25).
WARNING: Wireless communication equipment such as wireless home network devices, mobile phones, cordless
telephones, and their base stations, walkie-talkies, etc., can affect the MAXI MOVE and should be kept at least
2.34 km away from it. Cables from potentially strong sources of electromagnetic fields should not be placed near
the unit. See “Electromagnetic Compatibility” section for more details.
DIGITAL SCALE SPECIFICATIONS
Weight range
227 kg (500 lb)
Display resolution and type
0.1 kg (0.2 lb), liquid crystal display
Accuracy (in service)
2-50 kg ±100 g4-100 lb ±0.2 lb
Class III
50-200 kg ±200 g100-400 lb ±0.4 lb
200-227 kg ±300 g400-500 lb ±0.6 lb
OPERATION AND STORAGE CONDITIONS
Ambient temperature range
Operation: 5° to 40°C (+41° to +104°F) Storage: - 25° to 70°C (-13° to 158°F)
Relative humidity range
Operation: 15 to 93% (non condensing)
Storage: Up to 93%, (non condensing)
Atmospheric pressure range
Operation: 700 hPa to 1060 hPa (2000 m max) Storage: 500 hPa to 1060 hPa
WARNING: The equipment is not suitable in the presence of flammable
anesthetic mixture with air or oxygen, or with nitrous oxide. Using the MAXI MOVE in this environment might lead
to an explosion. The lift might create some spark internally and ignite the gaz.
RECYCLING
Battery
Sealed lead-acid, rechargeable, recycleable
Package
Cardboard recycleable
The Lift
Separated and recycled, according to the European Directive 2002/96/EG
(WEEE).
Technical Specifications