Product Description/Functions
14
001
.2
506
0.E
N
r
e
v.
1
1
System Failure Wind Down Facility
If electrical power fails completely due to battery
power loss or other electrical malfunction, the jib
can be lowered by first raising the red colored
emergency lowering lever found on the rear
section of the mast (See Fig. 9).
Fig. 9
Next, remove the exposed locking pin from its
location beneath the red colored emergency
lowering lever (see Fig. 10).
Fig. 10
Finally, using the lever as a crank, turn it in a
clockwise rotation (see Fig. 11). One full
clockwise rotation of the shaft lowers the mast jib
by 10 mm (3/8 in).
Fig. 11
If the wind down function must be used,
immediately remove the lift from use and contact
the ArjoHuntleigh Service Department or its
appointed distributor.
Automatic Cut-Out
This is not an operator control but a function built
into the lift’s electronics.
If the lift is inadvertently overloaded by trying to
raise or lower a load heavier than permitted, an
automatic “cut-out” function operates to prevent
the lift from raising a weight in excess of the safe
working load (SWL). This will stop the lift’s motion
automatically.
If this occurs, release the jib “up” button on the
handset or the control panel. Do not insist to raise
the load. Make sure that the MAXI MOVE
operates only within its safe working load.
Anti-Crush System
This is not an operator control but a function built
into the lift’s electronics.
Great care should be taken not to lower the
spreader bar, or stretcher onto the patient or any
other obstruction. If this should happen, the unit’s
“anti-crush” system will engage, stop the motor
and all downward movement will cease. If this
occurs, release the jib “down” button immediately
and press the jib “up” button to raise the jib until
the lift is clear. Then remove the obstruction.
Battery Indicator
The battery indicator for the MAXI MOVE is a
feature found on the control handset. Please refer
to the “Battery Charging” section for operating
procedures.
Sleep Mode
The MAXI MOVE is equipped with a power-
saving feature which places the machine in “sleep
mode” when not used for a short time. The unit is
put into sleep mode in two stages:
1) After two minutes of inactivity (where no
buttons are pressed on either the control
handset or control panel), the handset’s
display will go into sleep mode. The display
can be taken out of sleep mode by pressing
any button on the handset or the control panel.
There will be a three second delay after which
the unit is fully ready for use.
2) After six minutes of inactivity, the entire unit will
placed into sleep mode and will restart only
when a button is pressed either on the control
handset or control panel. There will be a three
second delay after which the unit is fully ready
for use.
WARNING: In order to prevent a fall risk and
injuries, if the mast is in a high position and the
wind-down function is used, ensure that suitable
and safe measures are taken to gain access to
the top cover.