Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test
IEC 60601-1-2 test level
Compliance level
Electromagnetic
environment – guidance
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT; 1 cycle
and
70% UT; 25/30 cycles
Single phase: at 0°
0% UT; 250/300 cycle
0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT; 1 cycle
and
70% UT; 25/30 cycles
Single phase: at 0°
0% UT; 250/300 cycle
NOTE -
U
T
is the AC mains voltage prior to application of the test level.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
product is used exceeds the applicable RF compliance level above, the product should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.