52
Thermometer
Range
10-50 °C
Accuracy
+/-1 °C
Resolution
0.1 °C
Operating Conditions
Ambient temperature
5 °C – 35 °C
Relative humidity
30% - 90% Rh
Atmospheric pressure
800 hPa to 1060 hPa
Transport and Storage
Ambient temperature
-20 °C to 70 °C
Relative humidity
10% - 80% at 20 °C
Atmospheric pressure
500 hPa to 1100 hPa
Mains Disconnection Device
Permanently installed in wall. visible and accessible at all times
Mains transient voltage
4 kV*
Creepage distance
3 mm*
Air clearance
1.8 mm*
Direction of movement
Up and down, right and left, clockwise and
counterclockwise
* If local regulations are higher, they must be met.
Water Connections
Cold/hot water connection
G 3/4” male (customers connection)
Maximum static pressure
6 bar (600 kPa, 87 psi)
Minimum operating pressure
1 bar (100 kPa, 15 psi)
Minimum drainage capacity
100 l/minute
Temperature - Cold water supply
5 - 20 °C
Temperature - Hot water supply
For WRAS approval the inlet water must be within:
For TMV3 approval the inlet water must be within:
55 - 60 °C
55 - 65 °C
The system is tested up to 80 °C