42
Disinfect (step 26-40)
26 Press the disinfection button on the control panel for
5 s. The LED will light up green
For bathtubs with Air Spa system:
Approximate time for the automatic disinfection is 3 to 4
minutes. 10 seconds after the start, the disinfection spray
handle is activated for 2 minutes before it shuts off.
27 Spray disinfectant into the drain.
28 Close the drainage plug.
29 Spray disinfectant on all internal
Malibu/Sovereign
bathtub
surfaces, lifting arm, transfer chair and the loose parts and
accessories at the bottom of the bathtub. Turn accessories
and spray on other side. Pay attention to parts and knobs
attached to the
Malibu/Sovereign
bathtub surface such as
drain control knob/surface overflow outlet.
30 Soak a cloth in disinfectant.
31 Wipe all other contact areas such as the outside of the
Malibu/Sovereign
bathtub (particularly behind the head),
metal sensory knobs, hand control and transfer chassis
(optional) with the cloth.
32 Allow an appropriate contact time of the disinfectant
according to the disinfectant IFU.
33
For bathtubs with Air Spa system:
Disinfection cycle will turn off automatically.
For bathtubs without Air Spa system:
Press the Disinfection button on the control panel to end
the disinfection process.
34 Put the disinfectant spray handle back.
35 Open the drainage plug.
36 Rinse all parts clean with water having a temperature of
approximately 25 °C using the shower handle to remove
the disinfectant.
37 Wet a new cloth with water and remove all traces of
disinfectant on contact areas. The cloth should occasionally
be rinsed in running water when removing disinfectant.
38 Place the shower handle in the holder.
39 Leave to air dry.
40 Attach armrests and seat. Make sure the seat is attached.
Attach the neck pillow, safety belt and rest of the
accessories.